SMFT-1000
Calibration Manual
18
R
INS
Insulation Resistance Method II (No Current)
To v e r i f y :
1. Set all equipment and the DUT to standby mode.
2. Connect shielded test leads from DUT red connector to the 532xA LO Port and DUT green
connector to the 532xA HI Port.
3. Disconnect all other test leads. See Figure 7.
Figure 7. R
INS
Insulation Resistance Method II (No Current) Connections
4. Turn the rotary switch on the DUT to R
INS
.
5. Push
to select R
INS
II mode.
6. Set the 532xA calibrator to output the first resistance test point and set the voltage output
of the DUT to the first voltage setting (see Ta b le 7 ).
A
N
A
L
YZE
R
Fluke 5320A
Fluke SMFT-1000
LO
METER
,
L
,
L
+/-