Multifunction PV Analyzer
Performance Tests
27
8. Push OPER on the 532xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
9. Push and hold
on the DUT until it beeps (~1 second).
Wait for the DUT to finish the measurement. Record the V
SPD
indication on the DUT in the
test record. Also record the maximum voltage measurement indication on the 532xA
calibrator in the test record. Multiply the resistance output of the 532xA calibrator (in
Ohms) by 0.001 A and use this as the reference voltage value to compare V
SPD
.
Note
Ignore the red X or green
. They do not indicate a pass or fail for this calibration.
10. Set the 532xA calibrator to output the next resistance test point (see Ta b l e 1 0 ).
11. Repeat steps 8 through 10 for all test points.
12. Set the 532xA calibrator to standby mode.
AC/DC Voltage
To veri f y :
1. Set all equipment and the DUT to standby mode.
2. Connect the low-thermal EMF test leads from DUT red connector to the 552xA NORMAL
(red) port and DUT blue connector to the 552xA black port below the NORMAL port.
3. Disconnect all other test leads. See Figure 11.
Figure 11. AD/DC Voltage Connections
4. Turn the rotary switch on the DUT to P
AC/DC
.
POWER
0
1
23
456
7
89
ENTER
M
k
m
VHz
FIELD
EDIT
/
+
F
OPR EARTH SCOPEEXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
F
n
p
p
W
dBm sec
C
A
MULT
x
DIV
MODES
MORE
STBY
5522A
CALIBRATOR
A
N
A
LYZE R
Fluke 5522A
Fluke SMFT-1000

Other manuals for Fluke SMFT-1000

Related product manuals