SMFT-1000
Calibration Manual
28
5. Push
until Voltage shows in highlight on the left-side menu.
6. Set the 552xA calibrator to output the first voltage test point (see Ta bl e 1 1 ).
7. Push OPER on the 552xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
Table 11. AC/DC Voltage Test Points
Range Frequency [Hz]
Te s t Po i n t s
[V]
Display Readings [V]
Lower Upper
0 V to 99.9 V
0
5.0 4.7 5.3
50.0 48.6 51.5
75.0 72.9 77.1
97.0 94.4 99.6
100 V to 1000 V
105 100 110
120 115 125
400 388 412
970 944 996
5 V to 99.9 V
50
5.0 4.7 5.3
50.0 48.6 51.5
75.0 72.9 77.1
97.0 94.4 99.6
100 V to 700 V
105 100 110
250 242 258
400 388 412
680 661 699
5 V to 99.9 V
60
5.0 4.7 5.3
50.0 48.6 51.5
75.0 72.9 77.1
97.0 94.4 99.6
100 V to 700 V
105 100 110
250 242 258
400 388 412
680 661 699