Multifunction PV Analyzer
Performance Tests
9
4. Set the 552xA calibrator to output the first voltage test point (see Ta bl e 3 ).
5. Push OPER on the 552xA to set the operate mode, if necessary, and wait for the calibrator
to stabilize.
6. Record the indication from the DUT in the test record.
7. Set the 552xA calibrator to output the next voltage test point (see Ta b l e 3 ).
8. Repeat steps 5 through 7 for all test points.
9. Set the 552xA calibrator to standby mode.
Tab le 3 . V
OC
Test Po int s
Range
Te s t Po i n t s
[V]
Display Reading [V]
Lower Limit Upper Limit
5 V to 99.9 V
5.0 4.8 5.2
50.0 49.6 50.5
80.0 79.4 80.6
99.2 98.5 99.9
100 V to 1000 V
103 100 106
500 496 505
990 983 997