SMFT-1000
Calibration Manual
32
Descending
0 A to 100 A
13.92
25-turn Coil
98.0 96.0 100.0
1 3 75.0 73.4 76.6
2 2 50.0 48.8 51.3
33
Single Wire
3.0 2.5 3.5
4 1 1.0 0.5 1.5
4 -1 -1.0 -1.5 -0.5
3 -3 -3.0 -3.5 -2.5
2-2
25-turn Coil
-50.0 -51.3 -48.8
1 -3 -75.0 -76.6 -73.4
1 -3.92 -98.0 -100.0 -96.0
AC Current
0 A to 100 A
41
Single Wire
1.0 0.5 1.5
3 3 3.0 2.5 3.5
22
25-turn Coil
50.0 48.8 51.3
1 3.4 85.0 83.2 86.8
1 3.92 98.0 96.0 100.0
41
Single Wire
1.0 0.5 1.5
3 3 3.0 5.2 3.5
22
25-turn Coil
50.0 48.8 51.3
1 3.4 85.0 83.2 86.8
1 3.92 98.0 96.0 100.0
Maximum DC
Hysteresis
Absolute maximum difference between like ascending and
descending DC test points
Table 13. AC/DC Current (Clamp only) Test Points (cont.)
Range Setup
552xA
Current
Output [A]
552xA Config.
Te s t
Points [A]
Display Readings
[A]
Lower Upper