• No cracks or crevices in, or around, the chamber walls to harbor
hidden or hard-to-reach bacterial growth.
• Stainless steel shelves, shelf channels and duct sheets which are easily
removed without tools for cleaning and disinfecting the entire interior.
• A blower wheel which is disposable and easily replaced.
• An improved water jacket drain and fill system.
Careful consideration has been given to the importance of product protection through
the addition of:
• An audible Add Water alarm and pilot light to alert the operator when
the water level in the water jacket has become too low for efficient
operation.
• Audible and visual CO
2
alarms which activate when the CO
2
content
in the chamber deviates 1% above or below the CO
2
control setpoint.
• An Over temperature Alarm which activates when the chamber temperature rises
above the control setpoint.
• An Add Water and Over temperature silence with 30 minute ring back.
• Remote alarm contacts for Over Temperature, Add Water, Power Failure and CO
2
deviation.