Forma Scientific 3158 - Complete CO2 Control Calibration Procedure

Forma Scientific 3158
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Forma Scientific, Inc.
7-5
2. Adjust the CO
2
setpoint to 0% by turning the CO
2
set screw on the control
module. (Refer to Figure 7-1) This will prevent CO
2
being injected while the
chamber is being calibrated.
3. Turn the CO
2
zero adjust screw on the control panel to match the digital
display with the reading of the test instrument.
4. Reset the CO
2
setting to the desired setpoint.
5. Open the doors to the chamber for 10 seconds to lower the CO
2
content in the
chamber.
6. Allow the incubator to stabilize for at least 30 minutes with both doors
closed.
7. Re-check the CO
2
concentration in the chamber with the test instrument. The
CO
2
should now be calibrated.
b. Complete CO
2
Control Calibration
Typical conditions for a complete calibration of the CO
2
control are:
After the CO
2
controller is replaced.
After the CO
2
sensor is replaced.
If there is reason to believe that the controller has previously been calibrated
incorrectly.
If the CO
2
concentration when checked by a Fyrite or other test instrument differs
by more than ± 1% from the display.
If laboratory procedures call for a periodic complete calibration.
Tools Needed:
Calibration screwdriver (provided).
Circuit board extender card (provided).
FYRITE CO
2
Analyzer.
1. Turn off all electric power to the incubator.