5-9
b. Over temp Alarm and Pilot Light
The over temperature audible alarm and pilot light will come on in the event of an over
temp condition. The audible alarm can be silenced for approximately 30 minutes by
pressing the SET/SILENCE button. The audible alarm will ringback every 30 minutes for
as long as the over temp condition exists.
c. Add Water Pilot Light and Audible Alarm
The ADD WATER audible alarm and pilot light are activated whenever the water level
in the water jacket is low. The audible alarm can be silenced for approximately 30
minutes by pressing the SET/SILENCE button. The audible alarm will ringback every 30
minutes for as long as the low water condition exists. (See Sections 4.10).
Temperature Monitor and Alarm Control Panel
Figure 5-6
5.7 REMOTE ALARM RELAY SYSTEM
Alarm states and alarm conditions are listed on Table 5-1. When the relay is off it is
in the alarm state.