Forma Scientific, Inc.
8-3
Inner Door 1/4" (0.62cm) Fully tempered safety glass with cam
action latch
Inner Door Gasket Non-porous feather edge silicone
Outer Door Gasket Four sided molded magnetic vinyl
Finish Powder coated for a durable, easily
maintained surface
Electrical
3154/3158 90-130 Volts 50/60HZ, 1PH, 4.2 FLA
(3164) 180-260 Volts 50/60HZ, 1PH, 2.6 FLA
3326/3354 90-130 Volts 50/60HZ, 1PH, 8.0 FLA
(3336) 180-260 Volts 50/60HZ, 1PH, 3.8 FLA
Circuit Breaker 5 Amps/chamber
Power Switch 2 Pole
Line Cord 8 ft.
Data Output 10mv/unit temperature & CO
2
(to Recorder output
connector)
Accessory Outlet 120V (75 Watts max.)
Remote Alarm Contacts Overtemperature, add water, power
failure, CO
2
deviation
Certification
CSA Standard C22.2 No. 151
UL Standard No. 1262 (except 3164/3336)
Dimensions
Exterior (Single) 24.37"W x 42.75"H x 23.31"F-B
(61.91 cm x 108.59 cm x 59.22 cm)
Interior 19.75"W x 26.75"H x 18.75"F-B
(50.17 cm x 67.95 cm x 47.63 cm)
Exterior (Stacked) 24.37"W x 83.00"H x 23.31"F-B
(61.89 cm x 210.82 cm x 59.20 cm)