IMPORTANT: Control solution results should fall within the control
solution range printed on the test strip vial. Do not use this range
when testing blood glucose.
English
English
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
23
What It Means What To Do
2. The te st
does not
start after
applying
the blood
drop.
• Blood drop is too small 1. Add blood to the same sample area within
60 seconds, or repeat the test using a new
test strip and a larger blood drop.
• Defective test strip
• Sample applied after
meter turns o
1. Repeat the test using a new test strip. Wait for
and to appear before applying blood or
control solution.
2. If problem persists, contact Customer Care.
• Defective meter 1. Contact Customer Care.
23-24ART22813-003_rev-A.indd
FreeStyle® Control Solution
Purpose
FreeStyle control solution is a red liquid that contains a xed amount of glucose.
The purpose of doing a control solution test is:
• To practice testing without using your own blood
• To conrm that the meter and test strips work together properly
• When you are unsure of your blood glucose test results
Contact Customer Care for information on how to obtain control solution.
Control Solution Information
• Use only FreeStyle control solution with the meter.
• Replace the cap securely on the bottle immediately after use.
• Control solution results are accurate only between 59° – 104° F (15° – 40° C).
• Do not add water or other liquid to control solution.
• There are three levels of control solution available (Low, Normal, High).
For information on how to obtain control solution, contact Customer Care.