IMPORTANT: Do not use control solution past expiration date. Discard
control solution 3 months after opening or on the expiration date
printed on the bottle whichever comes rst. (Example: open April 15,
discard July 15; write the discard date on the side of the bottle.)
Top
Printed side up
Sample areas
English
English
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
25
Out of Range Control Solution Results
• Repeat the test if control solution results are outside the range printed on
the test strip vial.
• Stop using the meter if control solution results are consistently outside the
range printed on the test strip vial. Contact Customer Care.
Causes of out-of-range results include:
• Expired or bad control solution
• Expired or bad test strip
• Testing error
• Watered-down control solution
• Meter malfunction
• Test strip deterioration
• Control solution test done outside 59° – 104° F (15° – 40° C)
ART22813-003_rev-A.indd 25-26
Performing a Control Solution Test
1.
Check control solution expiration date.
2. Remove test strip from vial.
3. Insert test strip to turn meter on.
Note: The meter turns o after 2 minutes of inactivity.
Remove and reinsert the unused test strip to restart
the meter.
If the meter does not turn on go to Troubleshooting section.