Consult instructions for use For in vitro diagnostic medical device
Temperature limitation Catalog number
Manufacturer Recycle
Batch code Serial number
Use by Do not reuse
The battery in this product should be removed and disposed in accordance with local
regulations for separate collection of spent batteries.
Definition of Symbols
English
English
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
Specifications (continued)
41
Operating
temperature
Power source
Result range
Sample
Sample size
Size
Weight
Meter cleaning and
disinfection
Lancing device cleaning
and disinfection
40° to 104° F (4° to 40° C)
One CR 2032, 3V lithium battery, replaceable
20 to 500 mg/dL
Fresh capillary and venous whole blood samples
0.3 microliter (300 nanoliters)
1.57 in (w) x 2.9 in (l) x 0.65 in (d) 4.0 cm (w) x 7.4 cm (l) x 1.7 cm (d)
1.0 to 1.1 oz. (28.3 to 31.2 g) including battery
The meter has a use life of 5 years, which is 522 cleaning and 522
disinfection cycles (2 cycles per week for 5 years).
The lancing device has a use life of 2 years, which is 210 cleaning
and 210 disinfection cycles (2 cycles per week for 2 years).