1. With the meter o, press and hold
m
for 3 seconds until appears.
2. Press
m
to bypass setting the optional reminder alarms. You may choose to
set these later (see Setting Reminder Alarms section).
4
Atkins SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S: Fingerstick glucose determination in shock. Ann Int Med 114: 1020-
1024. 1991.
5
Sandler M, Low-Beer T: Misleading capillary glucose measurements. Practical Diabetes 7: 210. 1990.
6
Wickham NWR, Achar KN, Cove DH: Unreliability of capillary blood glucose in peripheral vascular disease. Practical Diabetes 3: 100. 1986.
Setting Up the Meter
English
English
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
Limitations:
• For in-vitro diagnostic use only (external use only).
• Not for diagnosis of or screening for diabetes.
• Not for testing of neonates.
• The test strips are for single use only. Do not reuse test strips.
• The system can give false results greater than 10,000 feet (3,048 meters) above sea level.
• The system can give false results outside the hematocrit range: 15% to 65%.
• Individuals experiencing a hypoxia state may yield falsely low results.
• In situations of decreased peripheral blood ow, capillary blood testing may not be
appropriate as it may not represent the true physiological state. Examples would include but
are not limited to: severe dehydration caused by diabetic ketoacidosis or the hyperosmolar
non-ketotic state, hypotension, shock, or peripheral vascular disease.
4, 5, 6
• This device should not be used to test critically ill patients.
• Alternative site testing (AST) should not be used to calibrate continuous glucose monitors
(CGMs) nor for use in insulin dose calculations.
• AST testing should only be done during steady-state times (when glucose is not changing rapidly).
• For single patient use at home by a lay user.
• Severe dehydration or excessive water loss may cause false low results. If you believe you are
suering from severe dehydration, consult your physician immediately.
• Cholesterol up to 500 mg/dL or triglycerides up to 3,000 mg/dL do not signicantly aect test
results. However, glucose values in specimens beyond these levels, should be interpreted with
caution.
Purpose
The purpose of this section is to set the meter sound volume, time and date.
3
4
ART22813-003_rev-A.indd 3-4 1:28 PM11/9/16

Table of Contents

Other manuals for Freestyle Lite