SS1807-N002 Instruction of System - 36/106
Menu item Setting Default Description Remarks
R-gain
0 to 960
(increments
of 2 steps)
The displayed value cannot be changed.
The current gain values of the connected
Endoscope is displayed. To execute
automatic adjustment, select "Execute Auto
Adjust"
G-gain
0 to 960
(increments
of 2 steps)
B-gain
0 to 960
(increments
of 2 steps)
Execute Auto Adjust
When this item is selected and the Enter key
is pressed, automatic adjustment is executed
and the values before adjustment and those
after adjustment are displayed
Cancel Adjustment
The values after automatic adjustment are
returned to those before automatic
adjustment.
Reset to Factory Default
R-gain, G-gain, B-gain values are reset to
the factory defaults.
Execute writing
When this item is selected and then the
Enter key is pressed, the scope color
adjustment values are overwritten onto the
EEPROM.
[Note] When the adjustment values are
written onto the EEPROM, those
values are used for subsequent
operations.
[Note] When “Save and Exit” is selected
without selecting "Execute Writing" ,
the color adjustment value is applied
only to the current examination.
Save and
Exit
When this item is selected and then the Enter key is pressed, the current
settings are finalized and operation returns to the System Configuration
screen.
[Note] When the ESC key is pressed without selecting “Save and Exit”,
the current setting is canceled and operation returns to the System
Configuration screen.