SS1807-N002 Instruction of System - 37/106
< Scope Info >
Information on the connected Endoscope is displayed (Display only).
Displayed items vary depending on the version of the scope CPU on the connected
Endoscope and scope series.
A: 530 series scope (with the scope CPU Ver.2.0 or later). 550 series scope.
B: 530 series scope (with the scope CPU earlier than Ver.2.0).
Information Value Description A B
Type (16characters)
The type of the connected Endoscope Displayed
Not
displayed
Serial (16characters)
The serial number of the connected
Endoscope
Displayed
[Note]
Not
displayed
ID (3characters) The scope ID of the connected Endoscope Displayed Displayed
Number of
Operation
Times
(10characters)
The number of operation times of the
connected Endoscope
Displayed
Not
displayed
Repair
History
(6characters) The code of the repair history Displayed
Not
displayed
Repaired
Date
(4, 2 and 4
characters)
Date of repair (year/month/day) Displayed
Not
displayed
Factory
Shipment
Date
(4, 2 and 4
characters)
Factory Shipment Date (year/month/day)of
the Endoscope
Displayed
[Note]
Not
displayed
[Note] If the scope data supporting the EP-6000 is not written in the connected Endoscope, the
scope data possessed by the processor is displayed in place of the scope data
possessed by the Endoscope.
Accordingly, the displayed data does not match with the scope data of the connected
Endoscope.