ActionStep
Connect a suitable sample loop to the injection valve ports LoopF (fill) and
LoopE (empty).
Note:
Do not use a Superloop when cleaning the Injection valve.
2
Make sure there is waste tubing to the injection valve port W2 and W1 and
that they are secured in a waste container.
3
Fill a syringe with approximately 10 mL of an appropriate cleaning solution
(1 M NaOH or buffer solution). Connect the syringe to injection valve port
Syr, and inject the cleaning solution.
4
Fill a syringe with distilled water. Connect the syringe to injection valve port
Syr, and inject the distilled water.
5
ÄKTA go Operating Instructions 29360951 AC 95