ActionStep
Use a small flat-bladed screwdriver to prise open the fuse holder cover on
the power inlet.
2
Pull the fuse holder out of the mains connector panel by hand.3
Remove the fuses from the fuse holder.4
Fit new fuses of the same size and rating
1
.5
Replace the fuse holder in the power inlet.6
NOTICE
When replacing fuses, make sure that the fuse holder
is pushed fully into position.
Reconnect the power cord.7
1
Fuse rating: T4AL 250 V, 5 × 20 mm
98 ÄKTA go Operating Instructions 29360951 AC