DescriptionPart
Rinsing solution outlet, to the rinsing solution bottle7
Outlet, to the flow restrictor8
Function
Each system pump module consists of two pump heads. The individual pump heads are
identical but actuated in opposite phase to each other. The two pistons and pump heads
work alternately to give a continuous, low pulsation, liquid delivery. The two system
pumps can be used individually or in combination, to generate isocratic or gradient
elution in purification methods for flow rates up to 600 mL/min. The pumps can also be
used in combination to generate dual flow in the flow range 600-1200 mL/min, for iso-
cratic runs and limited gradient runs.
36 ÄKTA pilot 600 User Manual 29274559 AA
3 Description of modules
3.3 Pumps