EasyManua.ls Logo

GE AKTA pilot 600 - Background Considerations

GE AKTA pilot 600
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.1.1 Background considerations
Chromatography techniques
Different chromatography techniques make different demands on sample application,
as summarized in the table below. These considerations can influence the choice of
sample application technique.
More information about chromatography techniques may be found in the Life Science
Handbooks, available for download from the Support pages on www.gelifesciences.com.
DemandsTechnique
A sharply defined sample band with constant sample
concentration throughout. Any broadening of the sample
band will be reflected in broadening of the elution peaks.
Size exclusion chro-
matography (SEC)
The sample volume applied to the column should ideally
not be more than 10% of the column volume. Precise
requirements will depend on the separation being per-
formed.
Sample is concentrated by adsorption on to the chro-
matography resin during sample application. The volume
and concentration of sample is not critical, provided that
the capacity of the column is not exceeded.
Adsorption chromatogra-
phy, (e.g., affinity or ion
exchange chromatogra-
phy)
Sample application options
Samples may be applied to the column from a system inlet (A inlet) or from a Superloop
150 SNAP mounted on the column valve.
DescriptionWhen?Option
The sample is pumped onto the
column from one of the A inlets
using system pump A. The sample
passes through the system pump,
flow restrictor and any other mod-
ules before the column valve in the
flow path.
Recommended for sam-
ple volumes larger than
50 mL and for column
performance tests where
sample is readily avail-
able.
Apply sample from
a system inlet
The sample volume that can be
applied from a system inlet is in
principle unlimited.
ÄKTA pilot 600 User Manual 29274559 AA 61
4 Operation
4.1 Sample application
4.1.1 Background considerations

Related product manuals