Air trap indications in the
Process Picture
The table below explains the four different states.
ExplanationSymbol
The air trap is by-passed, so air in the flow path is not cap-
tured.
No air has been detected by the air sensor.
The air trap is by-passed, so air in the flow path is not cap-
tured.
Air has been detected by the air sensor. If the air sensor
alarm is enabled, the run is paused.
The air trap is in-line, so air in the flow path is captured.
No air has been detected by the air sensor.
The air trap is in-line, so air in the flow path is captured.
Air has been detected by the air sensor. If the air sensor
alarm is enabled, the run is paused.
Note:
Check if the air trap has run out of liquid.
Make sure that the the air trap is filled before setting it in-line. The recommen-
dation is to always have some liquid in the air trap because in-line position is
default. For information on how to fill the air trap, see the ÄKTA pilot 600
Operating Instructions.
Note:
When to use
Some examples when to use the air trap are shown below.
•
to trap degassed bubbles released from cold solutions
•
to catch any air introduced to the flow path during sample loading, for example
when applying all sample using an external air sensor
At flow rates > 100 mL/min the air trap can always be in-line.
44 ÄKTA pilot 600 User Manual 29274559 AA
3 Description of modules
3.6 Air trap valve and air trap