Using dual pump flow in a
method
Follow the steps below to use dual pump flow in a UNICORN method. These steps must
be performed for each phase that uses dual pump flow.
ActionStep
Open the Text Instructions tab for the phase.1
Select the System flow instruction.2
In the Instruction Box pane of the Text Instructions tab, select Pumps:Dual
pump flow.
3
Enter your parameters. Verify in the graph shown in Gradient applications,
on page 81 that the combination of %B and flow is compatible, and click
Change.
Tip:
When Pumps:Dual pump flow is selected, press the F1 key to access informa-
tion about the parameters.
4
ÄKTA pilot 600 User Manual 29274559 AA 83
4 Operation
4.4 Dual pump flow