Before every patient
This check does not need to be done before the first case of the day
if the "Every day before your first patient" checklist was done.
Check that the necessary emergency equipment is available
and in good condition.
Check vaporizer installation:
Make sure that the top of each vaporizer is horizontal (not
on crooked).
Make sure each vaporizer is locked and cannot be
removed.
Make sure the alarms and indicators operate correctly
(Tec 6 Plus vaporizer).
Make sure that more than one vaporizer cannot be turned
on at the same time.
Make sure that the vaporizers are adequately filled.
Check that an adequate reserve O2 supply is available.
Check that the breathing circuit is correctly connected, not
damaged, and the breathing system contains sufficient
absorbent in the canister.
Select Checkout and perform a Circuit Leak test.
Check that the ventilator functions correctly:
Connect a test lung to the patient breathing circuit
connection.
Set the ventilator to VCV mode and the settings for TV to
400 ml, RR to 12, I:E to 1:2, Tpause to Off, and Pmax to
40 cmH2O.
Set the gas flow to the minimum settings.
Start a case.
Set the Bag/Vent switch to Vent.
Fill the bellows using O2 flush.
Check that mechanical ventilation starts. Check that the
bellows inflate and deflate. Check that the display shows
the correct ventilator data. Check that there are no
inappropriate alarms.
Make sure that the alarms function. See the "Alarm tests".
Set the appropriate controls and alarm limits for the case.
Note
4 Preoperative checkout
2076152-001 4-3

Table of Contents