EasyManua.ls Logo

GE Carestation 650 A1 - Ventilation Modes Factory Default Settings

GE Carestation 650 A1
268 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tinsp
Rise Rate
AB.91.041
1
2
3
T
I
1. Paw
2. PEEP
3. Time
Figure 11-14 • CPAP + PSV waveform
Ventilation modes factory default settings
The ventilation mode factory default settings for the ADULT case
default show in the following table. An * indicates the setting is not
used for the ventilation mode. See the "Super user mode" section for
all other case default values.
.
Setting Range Mode
VCV PCV PCV-VG SIMV VCV SIMV PCV SIMV
PCV-VG
PSVPro CPAP
+PSV
TV 20-1500 ml 500 * 500 500 * 500 * *
RR 4-100 bpm 12 12 12 * * * * *
RR 2-60 bpm * * * 12 12 12 12 *
Minimum RR 4-60 /min * * * * * * * 12
I:E 2:1 - 1:8 1:2 1:2 1:2 * * * * *
Tpause Off, 5-60% of
Tinsp
Off * * Off * * * *
PEEP Off, 4-30
cmH2O
Off Off Off Off Off Off Off Off
Pmax 12-100
cmH2O
40 40 40 40 40 40 40 40
Pinsp 5-60 cmH2O * 5 * * 5 * 5 5
Psupport Off, 2-40
cmH2O
* * * 2 2 2 2 2
Tinsp 0.2-5 seconds * * * 1.7 1.7 1.7 1.7 1.7
11 Specifications and theory of operation
2076152-001 11-35

Table of Contents

Related product manuals