Airway module specifications
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. The following modules can be used on
this system:
• Airway Gas Option: N-CAiO
• CARESCAPE
™
series: E-sCAiO and E-sCAiOV
Gas specifications for airway modules
.
Airway humidity (patient
spirometry)
10% RH to 100% RH
Sampling delay 3.0 seconds typical with a 3 m sampling line
Total system response time Less than 3.8 seconds with a 3 m sampling line
Warm-up time 1 minute for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
20 minutes for full specifications
Respiration rate 4 to 100 breaths/min
Diversion flow 120 +/- 20 ml/min
Airway pressure -20 cmH2O to 100 cmH2O
.
Accuracy during stable conditions
Ambient temperature: 10 to 40°C
Ambient pressure: 495 to 795 mmHg
Ambient humidity: 10 to 98% RH, non-condensing
Automatic compensation for ambient pressure
Full module accuracy for Respiration rate of 4 to 70 breaths/min
CO2 +/- (0.2 vol% + 2% of reading)
O2 +/- (1 vol% + 2% of reading)
N2O +/- (2 vol% + 2% of reading) between 0 and 85 vol%
+/- (2 vol% + 8% of reading for N2O between 85 and 100 vol%