Pneumatic connections
Use only medical grade gas supplies. Other types of gas
supplies may contain water, oil, or other contaminants
which could affect the operation of the pneumatic system.
The gas supplies provide gas to these devices through internal
connections:
Venturi suction regulator (optional)
Auxiliary O2 flowmeter (optional)
Auxiliary O2+Air (optional)
ACGO (optional)
Pipeline inlets
AC.24.134
The pipeline inlets accept up to three different gases: O2, Air, and
N2O. A fourth pipeline inlet is available for a secondary, backup low-
pressure O2 supply connection.
Scavenging
The AGSS port is located behind the breathing system. An adapter
may be necessary to interface to the scavenging connector.
See "Passive AGSS" and "Active AGSS" for more scavenging
information.