Breathing system specifications
Compressible volume Ventilator side 2006 ml
Bag side 500 ml
Reusable canister 1000 ml
Disposable canister 1000 ml
Absorbent Reusable canister 1370 ml
Disposable canister 1400 ml
Connections Auxiliary Common Gas Outlet: ISO 5356 type
connector on the front of the system (standard 22
mm OD or 15 mm ID conical friction fit connectors).
Breathing system
leakage
Less than or equal to 150 ml/min total at 3 kPa (30
cmH2O) (both in absorber mode and with canister
removed).
Breathing system
compliance
Volume of gas lost due to internal compliance of
filled reusable absorber canister (bag mode only):
1.74 ml/0.098 kPa (1 cmH2O)
52.4 ml/3 kPa (30 cmH2O)
Volume of gas lost due to internal compliance of
filled disposable absorber canister (bag mode only):
1.81 ml/0.098 kPa (1 cmH2O)
54.56 ml/3 kPa (30 cmH2O)
APL valve Approximately 0 to 70 cmH2O
Negative pressure
relief
Minimum 100 ml/min at -14 cmH2O
Mechanical
overpressure valve
115 +/- 10 cmH2O
Inspiratory breathing resistance in bag mode*
l/min kPa cmH2O
Absorber canister installed
5 0.030 0.30
30 0.172 1.72
60 0.456 4.56
Absorber canister removed
5 0.034 0.34
30 0.159 1.59
60 0.401 4.01
*Values include patient circuit tubing and Y-piece 0.065 kPa (0.65 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations affect
resistance.
Carestation
™
620/650/650c (A1)
11-12 2076152-001