Calibration menu
Access the Calibration menu from the Checkout menu or by
selecting the System Setup button. Follow the instructions on the
screen.
The Calibration menu is not available during a case.
See "Calibrating the airway module" in the "Airway modules" section
for information on calibrating the airway modules.
Flow and pressure calibration
1. Select Checkout or System Setup.
2. Select Calibration on the Checkout or System Setup menu.
3. Select Flow and Pressure.
4. Follow the instructions on the screen.
Circuit O2 cell
Weekly O2 cell calibration
Weekly O2 cell calibration will calibrate the O2 cell to 21% O2,
through exposure to room air.
1. Select Checkout or System Setup.
2. Select Calibration on the System Setup menu.
3. Select Weekly O2 Cell.
4. Follow the instructions on the screen.
Monthly O2 cell calibration
Monthly O2 cell calibration will calibrate the O2 cell to 21% O2,
through exposure to room air, and to 100% O2, through exposure to
100% O2 supply.
1. Make sure the patient connection is not plugged or there is no
patient tubing connected to the system.
2. (ACGO option only.) Set the ACGO switch to Circle.
3. (Aux O2+Air option only.) Set the Aux O2+Air switch to Circle.
4. Set the Bag/Vent switch to Vent.
5. Select Monthly O2 Cell.
6. Follow the instructions on the screen.
Note
9 User maintenance
2076152-001 9-11