Setting alarm volume
Set alarm volume loud enough to be heard over
surrounding noise. If the volume is not loud enough, the
user may not hear the alarm.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Alarm Volume to the desired value.
The alarm volume range is 1 to 5.
4. Select Close.
Setting apnea delay
Use the Apnea Delay setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Apnea Delay to the desired time.
The apnea time delay range is 10 to 30 seconds.
The apnea time settings are in 1 second increments.
4. Select Close.
Silencing leak audio alarms
Use the Leak Audio setting to silence audio alarms of small leaks.
1. Select Alarm Setup.
2. Select the Configure tab.
3. Set Leak Audio to Off.
The audio alarms for small leaks are silenced.
4. To turn audio alarms back on, set Leak Audio to On.
5. Select Close.
WARNING
3 Operation
2076152-001 3-23