Message Priority Cause Action
FiDes low Low FiDes is less than alarm limit. Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiEnf high Medium FiEnf is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiEnf low Low FiEnf is less than alarm limit. Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiHal high Medium FiHal is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiHal low Low FiHal is less than alarm limit. Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiIso high Medium FiIso is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiIso low Low FiIso is less than alarm limit. Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
FiO2 high Medium FiO2 is greater than high alarm limit. Check the O2 setting. Recalibrate the
O2 cell and the airway module.
FiO2 low High FiO2 is less than low alarm limit. Check O2 setting. Check for leaks or
blockages in the patient circuit.
FiSev high Medium FiSev is greater than alarm limit. The
alarm priority escalates to High if the
alarm is not resolved after 2 minutes
Set the alarm limits appropriately.
Check agent setting.
FiSev low Low FiSev is less than alarm limit. Check the fill level of the vaporizers. Set
alarm limit appropriately. Check agent
setting.
Gas monitoring
not available
Medium Airway module hardware failure. Replace airway module. Then, turn
power off and back on between cases
to clear the alarm and receive module
data.
Gas controls
failure, O2 only
High System communication failure. Contact an authorized service
representative.
Gas controls
failure, O2 only
Medium Frame interface board and power
management board failure.
Contact an authorized service
representative.
Increase low MV
limit
Medium Low MV limit is off in SIMV VCV, SIMV
PCV, SIMV PCV-VG, CPAP + PSV, or
PSVPro
modes.
Increase low MV alarm limit to improve
patient disconnection detection.
Inspiration
stopped
Medium High airway pressure. Check system for blockages.
Internal failure.
System may shut
down.
High Power controller software failure. Contact an authorized service
representative.
Internal failure.
System may shut
down.
Medium Power controller software failure. Contact an authorized service
representative.
Carestation
620/650/650c (A1)
7-8 2076152-001