9. Test the Ppeak low. Leak? alarm:
• Unblock the patient connection.
• Set the Bag/Vent switch to Vent.
• Set the tidal volume and total flow to minimum.
• Other alarms such as MVexp low can occur.
• Make sure that the Ppeak low. Leak? alarm occurs.
10. Set all alarm limits to approved clinical values.
Carestation
™
620/650/650c (A1)
7-16 2076152-001