4-12 170 Series Monitor Revision D
2003023-001
Setup Procedures: Customizing the Monitor
Table 4-2. Summary of User Setup Codes
Code
(UA Display)
Setting or Value
(Primary FHR Display)
Code Description
1 FHR Alarms 0 = off (disabled)
1 = on (enabled)
2 FHR High Alarm Limit 140–210 (BPM, in increments of 5 BPM)
3 FHR Low Alarm Limit 50–140 (BPM, in increments of 5 BPM)
4 FHR Alarm Volume 2–10
10 Minutes (time setting) 0–59 (minutes)
11 Hours (time setting) 0–23 (hours)
12 Day of Month (date setting) 1–31 (day)
13 Month (date setting) 1–12 (month)
14 Year (date setting) 00–99 (year)