EasyManuals Logo
Home>GE>Medical Equipment>Corometrics 170 Series

GE Corometrics 170 Series User Manual

GE Corometrics 170 Series
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Revision D 170 Series Monitor 6-5
2003023-001
Uterine Activity Monitoring: Intrauterine Pressure Monitoring (Internal Method)
Intrauterine Pressure Monitoring (Internal Method)
Methodology
A catheter inserted transcervically into the uterine cavity measures intrauterine
pressure. You can monitor using either a fluid-filled catheter or a transducer-tipped
catheter. The uterine activity value is displayed in mmHg from 0–100 and the trend
is continuously plotted on the bottom (or right) grid of the strip chart paper as a plain
black line.
Pressure exceeding 100 mmHg is printed as a straight line at
100 mmHg.
Why Zeroing is Important
When you zero the system, you are referencing the pressure to
0 mmHg while the system is open to air to ensure an absolute pressure measurement.

Table of Contents

Other manuals for GE Corometrics 170 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 170 Series and is the answer not in the manual?

GE Corometrics 170 Series Specifications

General IconGeneral
BrandGE
ModelCorometrics 170 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals