33 Druck ADTS 2XX User Manual
K0451 Issue No. 1
SOUR:RATE
Command Syntax
SOURce:RATE<parameter> <aim>
Parameter Type DISCRETE, REAL
Static Channels: SOUR:RATE <parameter>,<Ps1 Rate Aim>,<Ps2 Rate Aim>
Pitot Channel: SOUR:RATE <parameter>,<Pt Rate Aim>
Parameter Range: ALT | CAS | MACH | PS | PT | QC
Function: To command the rate of change of the selected parameter to a new aim
value.
Conditions: Must be in control mode.
Query Syntax
SOURce:RATE? <parameter>
Parameter Type DISCRETE
Parameter Range: ALT | CAS | MACH | PS | PT | QC
Returned Static Channels: <Ps1 Aim>,<Ps2 Aim>
Returned Pitot Channel <Pt Aim>, *
Function: To query last rate command.
Conditions: None.