Druck ADTS 2XX User Manual 34
K0451 Issue No. 1
SOUR:STAT
Command Syntax
SOURce:STATe <state>
Parameter Type DISCRETE
Parameter Range: CONTROL | MEASURE | ON | OFF
Function: To change the mode of the pressure controllers.
CONTROL or ON to go to control mode.
MEASURE or OFF to go to measure mode.
Conditions: Must be in control mode.
Query Syntax
SOURce:STATe?
Parameter Type None
Parameter Range:
Returned Data: <state>
Returned Data Type: DISCRETE
Returned Data Range: ON | OFF
Function: To query state of pressure controllers.
Conditions: None