15 K0430 Issue No. 2
Pressure to Display (P-DISPLAY) Task
P-Display is a special application of Data Log. To use this mode, select Data Log from the
Store Mode menu as detailed on page 36. Connect the UUT to the instrument as shown
below and, if required, turn on and set-up Cal Mode (see page 11).
A - External pressure source B - Pressure regulator C - Dial gauge under test
•Press TASK and select P-DISPLAY. If required, use OUTPUT key to change
pressure units.
Set-up a data log file as detailed on page 30.
Note: TRIGGER field, automatically set to KEYPRESS, cannot be changed.
Apply a series of test pressures to the device under test. Enter displayed
reading at each pressure and log each point:
After logging final test point, terminate as follows:
A
B
PRESSURE INT bar
F2F1
DISPLAY
CHANGE
VALUE
TASK : P-DISPLAY
SNAPSHOT MODE
max 30V
DPI 615 IS
C
Taking Measurements