51 K0430 Issue No. 2
Current Output Range (24 mA)
Use the following procedure for calibrating the current output range.
(1) Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu as shown on page 43.
(2) Select CURRENT OUT range for calibration:
(3) Connect a digital milliammeter to the instrument as shown below.
Note: This procedure requires an external supply.
55mA
Vin
mA in
mA Sink
max 30V
max
mA
-
+
Voltage
source
Calibration