K0430 Issue No. 2 52
(4) Measure the 10% full-scale output current and enter the value measured on the
external milliammeter:
(5) Measure the 90% full-scale output current and enter the value measured on the
external milliammeter
(6) Press the ENTER key to accept the calibration. Press the EXIT key four times to
quit the CALIBRATION and SETUP modes.
(7) Verify the instrument’s output current calibration by setting up the loop test
currents as shown in Table 4 and checking the set values against the
milliammeter standard.
(8) Connect an external voltage source shown on page 51, press the OUTPUT key,
select mA Value:
Measure 90% FS - output
Press ENTER key when
applied value is stable
Store Cal point?
YES NO
Enter Measured Current:
RUN
:
OUTPUT
PRESSURE
mA STEP
mA RAMP
mA VALUE
SELECT OUTPUT
INT
PRESSURE EXT
UNITS
TASK: ADVANCED
SNAPSHOT MODE
VOLTAGE
V
TASK: ADVANCED
SNAPSHOT MODE
Calibration