35 K0430 Issue No. 2
Recalling Data Files (DPI 615 instruments only)
Data or results files generated by running procedures are stored in the instrument’s data
log directory. To recall a data file to the display, make sure that DATA LOG is selected
from the SETUP menu and proceed as follows:
Use the cursor keys to select either the AS FOUND option or the AS LEFT option for
display. AS FOUND is the result of the first run of a procedure and AS LEFT is the result of
the last time the procedure was run.
Procedure data files can be displayed either as text (stored screens) or in graphical form.
To display as text, select the TEXT option (F1) from the directory and proceed as follows
from the File Summary menu. Select
AUTO STEP (F1) to automatically review each
screen at 1 second intervals or use the left or right cursor keys to manually review.
To transmit the selected logged data via the RS232 interface, connect the instrument to a
free COM port on a compatible PC, ensure that the instrument’s RS232 parameters match
those of the PC.
To display logged data in graphical form, on screen, select GRAPH (F2) from the directory
and proceed as follows:
Memory Operations Datalog