K0430 Issue No. 2 56
(6) Apply the full-scale pressure to the external transducer and store the full-scale
(FS) point.
(7) Release the applied pressure and disconnect the pressure reference. Press the
ENTER key to accept the calibration. Press the EXIT key once to return to the
calibration menu or four times to quit the CALIBRATION and SETUP modes.
(8) Check the calibration of the external transducer by applying the test pressures as
detailed in Table 1, page 42.
Add External Sensor
To add an external sensor to the directory of external sensors, proceed as follows:
Connect the required external transducer to the EXT TRANSDUCER socket
located on the rear of the instrument.
Digitally compensated transducers will be downloaded into the directory as soon as the
F1 key is operated. Data for other types can be edited by selecting EDIT (F2). If the data
cannot be edited a warning message (INVALID ACTION) is displayed.
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
CALIBRATING PRESSURE EXT
RANGE 20000 kPa g
Apply nominal FS value
Press ENTER key when
applied value is stable
+
CHANGE VALUE