EasyManua.ls Logo

GE Druck DPI 615 IS

GE Druck DPI 615 IS
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
K0430 Issue No. 2 18
Relief Valve Test (REL VALVE) Task
To carry out a relief valve test, press TASK and select REL VALVE. Connect the output
pressure port of the instrument to an external system as shown below:
A - External pressure source (indicator only) B - Pressure regulator C - Relief valve under test
To change the pressure units, if required, press INPUT and select the required
units using the cursor keys.
Close the vent valve and, using the hand-pump or external pressure supply, apply
pressure to the relief valve under test.
When the relief valve operates, the maximum recorded pressure indicates the
operating point of the valve.
Record the test results.
Note: The STORE key can be used for this purpose. Use right cursor key initially,
followed by up/down keys to enter Snapshot text.
Open vent valve to release test pressure.
Note: If using external pressure supply, isolate supply before opening the vent
valve.
PRESSURE INT
bar
bar
RESET
MAXIMUM
MINIMUM
MAX/MIN
F2F1
A
B
C
PRESSURE INT
max 30V
TASK : RELVALVE
SNAPSHOT MODE
DPI 615 IS
3
Taking Measurements

Table of Contents