45 K0430 Issue No. 2
Voltage Input Range (5 Volts)
Use the following procedure for calibrating the internal 5 Volt range.
(1) Switch the instrument on, enter CALIBRATION mode and select INT RANGES from
the CALIBRATION menu shown on page 43.
(2) Select 5V range for calibration:
(3) Link the voltage input terminals with a short lead and enter the zero point:
(4) Remove the shorting link from the voltmeter terminals, connect the Vin terminals
of the instrument to a variable voltage source and connect a digital voltmeter
across the source:
SELECT CAL OPTION
SELECT CAL OPTION
SELECT RANGE
30 VOLTS
5 VOLTS
CALIBRATION
VOLTAGE IN
CURRENT IN
CURRENT OUT
PRESSURE INT
TEMPERATURE
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
Cal V
DVM
Calibration