EasyManua.ls Logo

GE Druck DPI 615 IS

GE Druck DPI 615 IS
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
v K0430 Issue No. 2
Specification
Safe working pressure
20 bar range (300 psi) 1.75 x full-scale
350 bar range (5000 psi) 1.2 x full-scale
400 bar range (6000 psi) 1.5 x full-scale
All other ranges 2 x full-scale
Accuracy
Combined non-linearity, hysteresis and repeatability
±70 mbar range (2 inHg) 0.05% F.S.
up to ±150 mbar (4.4 inHg) 0.05% span
200 mbar to 20 bar (3 psi to 300 psi) [Calibrator]): 0.025% F.S.
35 bar to 700 bar (500 psi to 10000 psi) [Indicator] 0.025% F.S.
70 bar to 400 bar (1000 psi to 6000 psi) [Hydraulic] 0.025% F.S.
Pressure Ranges
Refer to the pressure range matrix in the data sheet.
Temperature Effects
±0.004% of reading/°C (averaged over -10° to +40°C w.r.t. 20°C)
±0.002% of reading/°F (averaged over +14° to 104°F w.r.t. 68°F)
Power supply
Batteries 6 x 1.5 V C cells, alkaline (up to 60 hours nominal use at 20°C)
Voltage Inputs
Range: ±30V
Accuracy ±0.05% rdg, ±0.004% F.S.
Resolution 100µV max
Current Inputs
Range: ±55mA
Accuracy ±0.05% rdg, ±0.004% F.S.
Resolution 1µA max
Current sink
Range: 24mA
Accuracy ±0.05% rdg, ±0.01% F.S.
Resolution 1µA max
Display
Size: 60 x 60 mm (2.36” x 2.36”) LCD Graphics
Reading ±99999, update rate 2 readings/sec
Environment
Operating Temperature: -10°C to 50°C (+14°F to 122°F)
Calibrated Temperature: -10°C to 40°C (+14°F to 104°F)
Storage Temperature: -20°C to 60°C (-4°F to 140°F)
Calibration Temperature: 21°C ±2°C (70°F ±4°F)
Sealing
Sealed to IP54 (NEMA 4)
Physical
Size: 300 x 170 x 140 mm (11.8” x 6.7” x 5.5”)
Weight: 3 kg (6.6lb)

Table of Contents

Related product manuals