67 K0430 Issue No. 2
Typical Uploaded Data log File (DPI 610)
The following gives details of a typical data file upload.
FILENAME: TEST 5
DATE: 21/10/2006
TIME: 15:58
TRIGGER: KEYPRESS
AMBIENT TEMP : 24.1 C
NO. OF POINTS 11
RECORD TYPE 0
CURRENT mA PRESSURE INT bar
3.902 0.008 -0.65 %span 21/10/2006 15:58:55
6.076 2.311 1.42 %span 21/10/2006 16:00:03
7.598 4.404 0.47 %span 21/10/2006 16:00:57
8.085 5.023 0.41 %span 21/10/2006 16:01:45
9.949 7.249 0.94 %span 21/10/2006 16:02:42
13.002 11.300 -0.23 %span 21/10/2006 16:03:30
17.005 16.102 0.77 %span 21/10/2006 16:05:00
17.766 17.106 0.51 %span 21/10/2006 16:06:07
17.812 17.105 0.80 %span 21/10/2006 16:07:49
18.532 17.965 1.00 %span 21/10/2006 16:16:35
20.007 20.212 -1.02 %span 21/10/2006 16:18:23
The Record Type number contained in the header is a 16-bit field, with the individual bits
representing result formats as detailed below.
Record Type 1 P-Display
Record Type 2 Switch Test
Record Type 4 Leak Test
Record Type 8 Input channel: Min/Max ON
Record Type 120 Combination of all input process functions
Record Type 896 Combination of all output process functions
Record Type 1024 mA step mode
Record Type 2048 mA value mode
Record Type 4096 mA ramp mode
Record Type 8192 no input selected
Record Type 16384 no output selected