GE Druck DPI 880 - Switch test; UPM Pressure measurements

GE Druck DPI 880
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 - [EN] English
K405 Issue 3
Operation - Measure/supply Volts or mV
To measure/supply Volts or mV:
1. Connec
t the instrument (Figure 4/5) and, if necessary,
adjust the Set Up (Table 1).
2. Select the task from the task selection menu
(Table 2/3).
Note: Use the Dual Function area (
) to do two operations
at the same time. If the area
selection is not necessary,
set this area to off (). This saves the battery power.
3. If necessary, adjust the Settings (Table 4/5) and/or the
output values to the system (Table 7).
Figure 4: Example configuration - To measure/supply Volts
or mV (Area
)
Figure 5: Example configuration - To measure Volts (Dual
Function, area
)
Operation - Measure/supply Hz or pulses
To measure/supply Hz or pulses:
1. Connect
the instrument (Figure 6) and, if necessary,
adjust the Set Up (Table 1).
2. Select the task from the task selection menu (Table 2):
3. If necessary, adjust the Settings (Table 4/5) and/or the
output values to the system (Table 7).
Figure 6: Example configuration - To measure/supply Hz
or Pulses
For an input, the display shows the condition of the
frequency gate:
Operation - RTD/Ohms connections
In the examples that follow 2W, 3W, and 4W identify the 2,
3, and 4-wire connections for a RTD or resistance.
0 to ±30V OR
0 to ±120mV
Edit Settings
a) Measure
b)
Supp
l
y
0 to 12V OR
0 to 120mV
0 to ±30V
=
Gate open (measurement starts)
=
Gate closed (measurement is waiting for the next
rising edge of the cycle)
=
Fast cycle
SettingsEditSettings
a) Input b) Output