[EN] English - 13
K405 Issue 3
Calibration
Note: GE can provide a calibration service that is traceable
to international standards.
We recommend that you return the instrument to the
manufacturer or an approved service agent for
calibration.
If you use an alternative calibration facility, make sure that
it uses these standards.
Calibration - Before you start
To do an accurate calibration, you must have:
the
calibration equipment specified in Table 8.
a stable temperature environment: 70 ± 2°F (21 ± 1°C)
Table 8: Calibration equipment
Before you start the calibration, make sure that the time
and date on the instrument are correct (Table 1).
Selection sequence:
Task selection menu
Set Up (Table 1)
Calibration
Table 9: Calibration options
When you select a channel/function, the display shows the
applicable instructions to complete the calibration.
When the calibration is complete, select Calibration Due
and set the new calibration date for the instrument.
Function Calibration equipment
(ppm = parts per million)
mA OR
mA (Dual ... )
mA calibrator.
Accuracy - mA input/output: Table 10/11
Accuracy - mA (Dual Function): Table 10
mV OR
TC (mV)
mV calibrator.
Accuracy - mV input/output: Table 12/14
Accuracy - TC (mV): Table 20
Volts OR
Volts (Dual ... )
Volts calibrator.
Accuracy - Volts input/output: Table 13/ 15.
Accuracy - Volts (Dual Function): Table 13
Hz 1) Frequency meter
Total error: 7 ppm or better
Resolution: 8 digits (minimum)
2) Signal generator
IDOS UMM only. Refer to the user manual for the IDOS
UMM.
CJ - Standard RTD probe
Accuracy: 50 mK for 23 to 82.4°F (-5 to 28°C)
- Digital thermometer
Accuracy: 10 mK
RTD Ohms
- Standard 0Ω resistor
- *Standard resistor (Ω): 100, 200, 300
Tolerance: 50 ppm + 0.6 ppm/°C + 5 ppm/year
* Or an equivalent resistance simulator
- *Standard resistor (Ω): 400, 1k, 2k, 4k
Tolerance: 10 ppm + 0.6 ppm/°C + 5 ppm/year
RTD Ohms
An ohmmeter or an RTD measurement system
with the specified excitation currents (Table 19).
1
Display: Enter
Calibration PIN
23
Menu:
Select Channel
45
+
Factory PIN = 4321 (Table 9)
Select Function
(If applicable)
5a 5b
(Table 9)
Options Description
...
...
...
To calibrate the specified input/output:
... = mA, mV, Volts, Hz, RTD (Ohms), TC (mV)
IDOS UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
CJ To calibrate the cold junction channel.
mA (Dual ... ) To calibrate the mA (Dual Function) input.
Volts (Dual ... ) To calibrate the Volts (Dual Function) input.
Calibration Due: To set the date of the next
calibration for the instrument.
After the specified calibration date, there is a
warning message. There is a selection box to stop
the warning.
To change the calibration PIN (Personal
Identification Number).