18 - [EN] English
K405 Issue 3
Specification - Temperature ranges (RTD)
Specification - Resistance ranges (Ohms/RTD)
Specification - Frequency
cpm = counts/minute, cph = counts/hour
RTD type Standard Range °F Range °C Accuracy °F * Accuracy °C *
Pt50 (385) IEC 751 -328 to 1 562 -200 to 850 0.90 0.50
Pt100 (385) IEC 751 -328 to 1 562 -200 to 850 0.45 0.25
Pt200 (385) IEC 751 -328 to 1 562 -200 to 850 1.08 0.60
Pt500 (385) IEC 751 -328 to 1 562 -200 to 850 0.72 0.40
Pt1000 (385) IEC 751 -328 to 752 -200 to 400 0.36 0.20
D 100 (392) JIS 1604-1989 -328 to 1 202 -200 to 650 0.45 0.25
Ni 100 DIN 43760 -76 to 482 -60 to 250 0.36 0.20
Ni 120 MINCO 7-120 -112 to 500 -80 to 260 0.36 0.20
*Temperature coefficient:
14 to 50°F, 86 to 122°F = 0.0028% FS / °F
(-10 to 10°C, 30 to 50°C = 0.005% FS / °C)
Range (Ω) Excitation (mA) Accuracy (Ω)*
Measure Supply
0 to 400 0.10 to 0.5 - 0.15
0 to 400 0.50 to 3.0 0.10 0.10
400 to 1 500 0.10 to 0.8 0.50 0.50
1 500 to 3 200 0.05 to 0.4 1.00 1.00
3 200 to 4 000 0.05 to 0.3 1.30 1.30
*Temperature coefficient:
14 to 50°F, 86 to 122°F = 0.0028% FS / °F
(-10 to 10°C, 30 to 50°C = 0.005% FS / °C)
Range (Measure): Accuracy:
0 to 999.999 Hz
0 to 50.0000 kHz
cpm: 0 to 999 999
cph: 0 to 999 999
For all the ranges:
0.003% of reading + 2 counts
ppm = pulses/minute, pph = pulses/hour
Range (Supply): Accuracy:
0 to 999.99 Hz
0 to 50.000 kHz
ppm: 0 to 59 999
pph: 0 to 99 999
0.003% of reading + 0.0023 Hz
0.003% of reading + 0.0336 Hz
0.003% of reading + 0.138 cpm
0.003% of reading + 0.5 cph
Temperature coefficient
14 to 50°F, 86 to 122°F
(-10 to 10°C, 30 to 50°C)
0.0017% FS / °F
(0.003% FS / °C)
Output waveform Square, bipolar
Voltage input 0 to 30 V
Trigger level 0 to 12 V, Resolution: 0.1 V
Output amplitude 0.1 to 12 V dc ± 1% (≤ 10 mA)
0.1 to 12 V ac* ± 5% (≤ 10 mA)
*Peak to Peak