EasyManua.ls Logo

GE Druck DPI 880 - Measure;simulate an RTD or Ohms; Thermocouple (TC) connections

GE Druck DPI 880
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 - [EN] English
K405 Issue 3
Table 5: (Part of table) Menu options - Settings (Output)
Table 6: Additional data for Settings (Output):
Operation
This section gives examples of how to connect and use the
instrument. Before you start:
Read and understand the “Safety” section.
Do not use a damaged instrument.
Operation - Electrical connections
To prevent instrument errors, make sure that the electrical
c
onnec
tions (A1-item 10 and/or A2) are correct.
Operation - Communications port connections
Use the communications port (A1 - item 9) to attach an
IDOS Univ
ersal Measurement Module (UMM).
When you attach the cable from a UMM (Figure 13/14), the
instrument automatically changes the menus to give you
all the applicable options (Table 2/3).
Operation - Change the output values
When the output operation is set up (Table 5), use one of
these procedures to change the output values:
Table 7: Procedures to change the output
Example procedure (“Nudge” output):
Options
(If applicable)
Description
To select and set up values for the “Span Check”
output. Example output cycle:
This cycle repeats
automatically.
Additional data (Table 6): Select Settings ( )
% Step
To select and set up values for the “% Step” output.
Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings ( )
... Step
To select and set up values for the “Defined Step”
output. Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings ( )
To select and set up values for the “Ramp” output.
Example output cycle:
Auto Repeat - Optional
Additional data (Table 6): Select Settings ( )
Item Value
Span Check
Low (0%) Set the 0% value.
High (100%) Set the 100% value.
Dwell (d) Set the period (Hours:Minutes:Seconds) between
each change in value.
% Step Low (0%), High (100%), Dwell (d): As above.
Step Size (s) ...
%
Set the change in value for each step as a
percentage of the full-scale range (High - Low).
Defined Step Low (0%), High (100%), Dwell (d): As above.
Step Size (s) Set the change in value for each step.
Example: 1.000 mA steps.
Ramp Low (0%), High (100%), Dwell (d): As above.
Travel (t) Set the period (Hours:Minutes:Seconds) to go from
the Low (0%) value to the High (100%) value.
Auto Repeat If applicable, select this item to repeat a cycle
continuously.
The Help button (A3 - Item 25) shows a connection
diagram for the task selections you have set.
Output Procedure
Select Edit ( ) and/or use the buttons.
See the example below.
,
Select Start/Stop ( ) or use the buttons
to make the step changes manually.
Select Start/Stop ( ).
Display:
mA output
1 Edit 23Edit 4
+01.000
Sk1 = Edit
Sk2 = Settings
Display:
mA output
56
2.000
3.000
Increment = 1.000
(Table 5)

Related product manuals