IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
3-46 Chapter 3 - System Setup
PRELIMINARY
Figure 3-28 DICOM Storage Properties
3-8-4-6 DICOM SR
DICOM Structured Reporting (SR) is a standardized format for medical results. Venue Go supports
the specialized form for Echo Ultrasound (“TID 5200 Echocardiography Procedure Report”) and
Vascular Ultrasound (“TID 5100 Vascular Ultrasound Procedure Report”) for M&A results.
With the DICOM SR support, M&A for an exam can be sent at the end of the exam or when exported
from local archive. The destination can be either a server on the network (Storage SCP) or a removable
media (DICOM Media) depending on the DICOM dataflow selected.
“TID 5200 Echocardiography Procedure Report” is sent if the exam contains M&A from category
Cardiac or Pediatric (Heart). “TID 5100 Vascular Ultrasound Procedure Report” is sent if the exam
contains M&A from category Vascular or Abdominal. If the exam contains M&A from both Cardiac/
Pediatric (Heart) and Vascular/Abdominal categories, two SR documents are sent.
“TID 5200 Echocardiography Procedure Report” and “TID 5100 Vascular Ultrasound Procedure
Report” do not support all M&A results from Venue Go™ . They are limited to the following:
No unassigned measurement.
Refer to the Venue Go™ Reference manual for a complete list of supported parameters.
The following modes: 2D, M-mode, Color Flow, PW Doppler, CW Doppler, 3D and TDI.
Not Modified Simpson method or Bullet methods.
Refer to the Venue Go™ Reference manual for a complete list of supported methods.
Basic derivations (Average, Last, Min and Max), no references between the derived measurements
and the ones they were made from.
Wall Motion Scoring: individual segment scores only according to 16-segment model, no graded
Hypokinesis (only Hypokinesis is used).
DICOM SR must be activated for each DICOM device.
1) Tap Utility/Config on the Touch panel and log on as administrator.
2) Select the Connectivity category and Dataflow subgroup.
The Dataflow menu is displayed.