GE HEALTHCARE
DIRECTION 5162630, REVISION 3 LOGIQ™ S6 SERVICE MANUAL
8-52 Section 8-3 - Software Loading Procedures (R7.7.x or later with BEP4)
8-3-10-6 Regional Setup (For the system with language other than English ONLY) (cont’d)
3.) Select the Language tab, press Details, under Installed Services select the Russian or Greek
keyboard, under Default input language select Russian - Russian or Greek - Greek, press Apply,
Press OK.
4.) Select the Advanced tab, then select Russian or Greek in the Language for non-Unicode
programs pull-down menu. Press Apply. Answer Yes to use files already loaded on the hard disk,
then answer No to not reboot the system yet, press OK. Press Save and Exit the Utility screen.
5.) Reboot the system. When your system restarts, the system appears in the selected language.
6.) To switch between the English and Russian (or Greek) keyboard, press Alt+Shift to change the
keyboard to the Russian or Greek keyboard.
7.) Apply the changes by pressing Apply. Press OK TWICE
.
Figure 8-173 Advanced Tab
NOTICE
To have the settings take effect, you MUST turn off the system and turn it back on.