General Procedures and Functional Checks
4-36       LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-3-24 Tissue Velocity Imaging (TVI) Checks
4-3-24-1 Introduction
TVI calculates and color codes the velocities in tissue. The 
tissue velocity information is acquired by sampling of tissue 
Doppler velocity values at discrete points. The information is 
stored in a combined format with grey scale imaging during one 
or several cardiac cycles with high temporal resolution.
4-3-24-2 Preparations
•  Connect one of the probes, to the scanner’s left-most probe 
connector. 
• See 4-3-27 ‘Probe/Connectors Check’ on page 4-39 for info 
about connecting the probes.
For available probes, see 9-3-2 ‘Probe’ on page 9-5: 
•  Turn ON the scanner.
The 2D Mode window is displayed (default mode).
•  If needed, adjust the Display’s Brightness and Contrast 
setting. 
Press TVI.
Use the trackball (assigned function: Pos) to position the 
ROI frame over the area to be examined.
Press Scan Area. The instruction Size should be 
highlighted in the trackball status bar.
NOTE:  If the trackball control pointer is selected, press trackball to 
be able to select between Position and Size controls.
Use the trackball to adjust the dimension of the ROI.